USB-9211/9211A User Guide and Specifications
10
ni.com
Figure 8.  Input Circuitry for One Channel
If thermocouples are connected to the USB-9211/9211A, the gain and 
offset errors resulting from the source impedance of the thermocouples are 
negligible for most applications. Other voltage sources with a higher source 
impedance can introduce more significant errors. For more information 
about errors resulting from source impedance, refer to the 
 
section.
Determining Temperature Measurement Accuracy and 
Minimizing Errors
Temperature measurement errors depend in part on the thermocouple type, 
the temperature being measured, the accuracy of the thermocouple, and the 
cold-junction temperature.
Using the Autozero Channel
The USB-9211/9211A has an internal autozero channel for measuring the 
offset error. If the ambient temperature of the USB-9211/9211A is less than 
15 °C or more than 35 °C, use this channel to read the offset error. The 
USB-9211 always has this channel enabled. Use NI-DAQmx to configure 
the behavior of the autozero channel for the USB-9211A. Refer to the 
NI-DAQmx Help, accessible from Start»All Programs»National 
Instruments»NI-DAQ, for more information.