Справочник Пользователя для National Instruments 9211

Скачать
Страница из 19
© National Instruments Corporation
11
USB-9211/9211A User Guide and Specifications
Measurement Accuracy for the Different Types 
of Thermocouples
Figures 910, 11, 12, and 13 show the typical and maximum errors for the 
different thermocouple types when used with the USB-9211/9211A over 
the full temperature range. The figures also show the maximum error for 
the thermocouple types with the USB-9211/9211A at room temperature, 
15 to 35 °C. The figures account for gain errors, offset errors, differential 
and integral nonlinearity, quantization errors, noise errors, and isothermal 
errors. The figures do not account for the accuracy of the thermocouple 
itself.
Temperature gradients across the USB-9211/9211A terminals affect the 
cold-junction temperature accuracy. Refer to the 
 section for more information about 
temperature gradients.
Figure 9.  Type B Errors
200
400
600
800
1200
1600
1800
0
2
4
6
Max over Temp Range with Autozero
Max at Room Temp Range without Autozero
Typ over Temp Range with Autozero
Error (
°C)
1000
1400
Measured Temperature (
°C)