Справочник Пользователя для National Instruments NI 9219

Скачать
Страница из 36
© National Instruments Corp.
15
NI 9219 Operating Instructions
lead wires have the same resistance. A gain of 2x is applied to the 
voltage across the negative lead wire and the ADC uses this voltage 
as the negative reference to cancel the resistance error across the 
positive lead wire. Refer to Figure 8 for an illustration of the 
connections.
Figure 8.  Connections in 3-Wire RTD Mode
2-Wire Resistance and Quarter-Bridge Modes
In 2-Wire Resistance and Quarter-Bridge modes, connect the 
two ends of the resistor or gauge to the NI 9219 across the HI and 
LO terminals. These modes source a current, which varies based on 
the resistance of the load, between the HI and LO terminals. The 
resulting resistance is computed from the voltage measurement. 
2-Wire Resistance and Quarter-Bridge modes do not compensate 
LO
EX+
EX–
RTD
Wire
Wire
Wire
I
x2
ADC
I
I
NI 9219