Справочник Пользователя для National Instruments 5441
NI 5441 Specifications
4
ni.com
Accuracy
DC Accuracy
For the Low-Gain or High-Gain Amplifier Path:
±0.2% of Amplitude ± 0.05% of Offset ± 500 µV
(within
(within
±10 °C of self-calibration temperature)
±0.4% of Amplitude
± 0.05% of Offset ± 1 mV
(0 °C to 55 °C)
All paths are
calibrated for
amplitude and
gain errors. The
Low-Gain and
High-Gain
Amplifier Paths
also are
calibrated for
offset errors.
calibrated for
amplitude and
gain errors. The
Low-Gain and
High-Gain
Amplifier Paths
also are
calibrated for
offset errors.
For the Direct Path:
Gain Accuracy:
±0.2% (within ±10 °C of self-calibration
temperature)
Gain Accuracy:
±0.4% (0 °C to 55 °C)
DC Error:
±30 mV (0 °C to 55 °C)
AC Amplitude
Accuracy
Accuracy
±1.0% of Amplitude ± 1 mV
50 kHz sine
wave.
wave.
Output Characteristics
Output
Impedance
Impedance
50
Ω nominal or 75 Ω nominal, software-selectable.
—
Load
Impedance
Compensation
Impedance
Compensation
Output amplitude is compensated for user-specified load
impedances.
impedances.
—
Output
Coupling
Coupling
DC
—
Output
Enable
Enable
Software-selectable. When disabled, CH 0 out is terminated
with a 1 W resistor with a value equal to the selected output
impedance.
with a 1 W resistor with a value equal to the selected output
impedance.
—
Maximum
Output
Overload
Output
Overload
The CH 0 output can be connected to a 50
Ω, ±12 V
(
±8 V for the Direct Path) source without sustaining any
damage. No damage occurs if the CH 0 output is shorted to
ground indefinitely.
ground indefinitely.
—
Waveform
Summing
Summing
The CH 0 output supports waveform summing among
similar paths—specifically, the outputs of multiple NI 5441
signal generators can be connected together.
similar paths—specifically, the outputs of multiple NI 5441
signal generators can be connected together.
—
Specification
Value
Comments