Справочник Пользователя для National Instruments 5441
© National Instruments Corporation
5
NI 5441 Specifications
Frequency and Transient Response
Bandwidth
43 MHz
Measured at
–3 dB.
–3 dB.
DAC Digital
Interpolation
Filter
Interpolation
Filter
Software-selectable Finite Impulse Response (FIR) filter.
Available interpolation factors are 2, 4, or 8.
Available interpolation factors are 2, 4, or 8.
Refer to the
section for OSP
Interpolation.
Interpolation.
Analog
Filter
Filter
Software-selectable 7-pole elliptical filter for image
suppression.
suppression.
Available only
on Low-Gain
Amplifier and
High-Gain
Amplifier Paths.
on Low-Gain
Amplifier and
High-Gain
Amplifier Paths.
Passband
Flatness
Flatness
Path
—
Direct
Low-Gain
Amplifiers
High-Gain
Amplifiers
Amplifiers
+0.6 dB to –0.4 dB
100 Hz to 40 MHz
+0.5 dB to –1.0 dB
100 Hz to 20 MHz
+0.5 dB to –1.2 dB
100 Hz to 20 MHz
Pulse
Response
Response
Path
Analog Filter
and DAC
Interpolation
Filter disabled.
and DAC
Interpolation
Filter disabled.
Direct
Low-Gain
Amplifier
High-Gain
Amplifier
Rise/Fall Time
<5 ns
<8 ns
<10 ns
Aberration
<10%
<5%
<5%
Specification
Value
Comments