Справочник Пользователя для National Instruments pci-6110e-6111e

Скачать
Страница из 113
Chapter 4
Signal Connections
4-24
©
 National Instruments Corporation
Refer to Figures 4-8 and 4-9 for the relationship of STARTSCAN to the 
DAQ sequence.
As an input, the CONVERT* signal is configured in the edge-detection 
mode. You can select any PFI pin as the source for CONVERT* and 
configure the polarity selection for either rising or falling edge. The 
selected edge of the CONVERT* signal initiates an A/D conversion.
As an output, the CONVERT* signal reflects the actual convert pulse 
that is connected to the ADC. This is true even if the conversions are 
being externally generated by another PFI. The output is an active low 
pulse with a pulse width of 50 to 100 ns. This output is set to tri-state at 
startup.
Figures 4-18 and 4-19 show the input and output timing requirements 
for the CONVERT* signal.
Figure 4-18.  CONVERT* Input Signal Timing
Figure 4-19.  CONVERT* Output Signal Timing
The ADC switches to hold mode within 20 ns of the selected edge. This 
hold-mode delay time is a function of temperature and does not vary 
from one conversion to the next. 
Rising-edge
polarity
Falling-edge
polarity
t
w
t
w
= 10 ns minimum
t
w
t
w
= 50-100 ns