Chapter 4
Connecting Signals
© National Instruments Corporation
4-27
Figure 4-20 shows the timing for SCANCLK.
 
Figure 4-20.  SCANCLK Signal Timing
EXTSTROBE* Signal
EXTSTROBE* is an output-only signal that generates either a single pulse 
or a sequence of eight pulses in the hardware-strobe mode. An external 
device can use this signal to latch signals or to trigger events. In the 
single-pulse mode, software controls the level of EXTSTROBE*. A 10
µs 
and a 1.2
µs clock are available for generating a sequence of eight pulses in 
the hardware-strobe mode.
Note
EXTSTROBE* cannot be enabled through NI-DAQ.
Figure 4-21 shows the timing for the hardware-strobe mode 
EXTSTROBE* signal.
 
Figure 4-21.  EXTSTROBE* Signal Timing
Waveform Generation Timing Connections
The analog group defined for the NI PCI-6110/6111 is controlled by 
WFTRIG, UPDATE*, and UISOURCE. 
t
= 450 ns
t
= 50 to 100 ns
CONVERT*
SCANCLK
t
d
t
w
t
= 600 ns or 5 µs
V
OH
V
OL
t
w
t
w