Справочник Пользователя для CVRx BAROSTIMIPG1
BAROSTIM NEO LEGACY REFERENCE GUIDE
9-4
Incorrect severe lead angle from header
Incorrect placement of lead behind IPG
Figure 11: Incorrect placement of excess lead body
7. Tie the suture used to secure the IPG to the fascia.
8. If a subfascial pocket is used, close the fascia over the IPG.
9. Irrigate the pocket with an antibiotic solution.
10. Check the impedance of the system to assure the adequacy of the electrical connections.
11. Consider infiltrating the incisions with local anesthetic
12. Close the incisions per the surgeon’s usual practice.
Pre Discharge Device Testing
1. The patient should be seated or in a hospital bed at an angle of approximately 45°-90°. The patient should
sit with back and arms supported for at least 5 minutes before blood pressure measurements are taken.
Push the Edit and Test button for one of the Therapies.
Push the Edit and Test button for one of the Therapies.
2. Measure baseline hemodynamic and/or physiological values
3. Initiate CSL activation (typically at 2 mA and 65µs) by checking the box of the desired Pathway and
pressing Test Now.
4. It is recommended that the IPG be programmed to lower settings (Including Off) if:
• Concerning changes in hemodynamics occur or
• Problematic tissue stimulation is noted or
• Any other potentially hazardous patient responses are observed
• Problematic tissue stimulation is noted or
• Any other potentially hazardous patient responses are observed
5. Wait approximately 1 minute
6. Measure and Record the patient’s hemodynamic and/or physiological response.
7. Repeat steps 2 through 6 at increasing pulse amplitude to optimize settings.
8. Other parameters including, but not limited to, pulse width, frequency, and pathway (if applicable) can be
tested independently using a similar process if desired.