 User Manual, Version 1.09 
31 
Copyright © 2006 by Datastrip, Inc.  All rights reserved.   
Reproduction in whole or in part is prohibited. 
8.   Loading Software and Firmware 
8.1  Loading DSV2+ Software Updates 
The DsVerify2SoftwareUpdate application permits the user to download software updates from 
several source locations are as list below: 
  CAB files located on \Storage Card 2, \Storage Card, and USB thumb drives 
  CAB files loaded directly from the Datastrip corporate FTP server, (requires a Datastrip 
supplied) username and password that accesses a pre-determined account.   
  CAB files loaded from user/developer designated FTP servers, given a username and 
password to a valid account. 
8.1.1 
DSV2+ Software Update Procedure - Using CF card in external slot 
NOTE:  It is recommended that upgrades are done with the unit operating on the AC adapter.  
This upgrade process should take approximately 3.5 minutes to complete. 
  Power up unit. 
  Insert the provided CF Card containing the new OS cab file into the external CF slot. 
  When the device is finished booting, double tap My Computer and Select Storage Card / 
DSVII. 
  Double tap the DsVerify2SoftwareUpdate utility.   
  Select Server: Storage Card2 (you must scroll down to make selection) 
  Username and Password are not necessary for this procedure 
  Select most recent appropriate source file (Ex.  Src File: \DSVII_0195_sh4.CAB) 
  Click Update (This step will take approximately 2.5 minutes. 
  Click Reboot (The registry will be cleared prior to shutdown). 
  Power Up unit 
  The DSV2+ registration program will automatically open.  Select the button for the 
device that you are registering (DSVII-SC, DSVII-PA, or DSVII-SW).  Select [OK] and 
wait.  (This process will take from a few seconds). 
  Click [Ok] in the “Hurray!” Box when it appears, then close the application window. 
  Double tap the RegistrySaver program. 
  Double tap My Computer and Select Storage Card / DSVII. 
  Double tap the PPCalFt41utility and press the “Calibrate” button to initialize the 
fingerprint sensor.  This process may take up to a minute and then a “green” window with 
a number “1” should appear to indicate the sensor was successfully calibrated.  
  Double tap the DsVerifyFpCaptureTest application. In the lower right hand corner select 
“UPEK” and click “OK”.  DO NOT use the calibrate button on this screen for UPEK, 
however you can use the Scan function to test the fingerprint image. 
  If you are upgrading a scanner go to Section 8.2 
8.1.2 
DSV2+ Software Update Procedure – Using Datastrip FTP site 
NOTE: It is recommended that upgrades are done with the unit operating on the AC adapter.  
This upgrade process should take approximately 13 minutes to complete. 
  Power up unit. 
  When the device is finished booting double tap My Computer and Select Storage Card / 
DSVII. 
  Double tap the DsVerify2SoftwareUpdate utility 
  Select Server: datastrip.serveftp.net 
  Click Connect