Справочник Пользователя для Instrumentarium Dental PaloDEx Group Oy SM

Скачать
Страница из 50
6 Technical specifications
200150  rev  3
Instrumentarium Dental
33
6 Technical specifications
6.1
G
ENERAL
D
Manufacturer´s quality 
system:
ISO 9001, ISO 13485
Electrical and Mechanical 
safety:
Medical Equipment with 
respect to electrical 
shock, fire and 
mechanical hazards 
only in accordance with UL60601, 
CAN/CSA C22.2 no 601.1
According to IEC 60601-1, CE models 
marked according to the Medical 
Device Directive 93/42/EEC (Device 
classification: II b),
Mobile sensor unit that is IEEE 
802.11b  (WLAN) compliant
Patient connection (sensor):
BF-type
Sensor technology:
CMOS Active Pixel technology 
(CMOS APS)
IMAGE QUALITY
Resolution:
26.3 lp/mm (theoretical)
Grayscale:
4096 gray levels.
Holders:
A full set of color coded KerrHawe and 
Sigma M sensor holders, with easy 
instructions, for comfortable and 
accurate positioning.
SOFTWARE MAIN 
FEATURES
Operating system:
Windows 2000 or XP (SP2)
Operating software:
CliniView version 6.4 or later
Full Mouth Series:
FMS, user defined
File sizes:
3,17 MB, uncompressed file,Sensor 
size1 4,96 MB, uncompressed 
file,Sensor size2
X-ray shield in the sensor:
Lead (Pb)
6.2
D
IMENSIONS
Outer dimensions:
Sensor unit:
102 x 60 x 22 mm
Sensor size 1:
26 x 37 mm
Sensor size 2:
31 x 44 mm
Active image area:
Sensor size 1:
30.02 x 19,95 mm
2
Sensor size 2:
36,48 x 25,84 mm
2
6 Technical specifications
200150  rev  3
Instrumentarium Dental
33
6 Technical specifications
6.1
G
ENERAL
D
Manufacturer´s quality 
system:
ISO 9001, ISO 13485
Electrical and Mechanical 
safety:
Medical Equipment with 
respect to electrical 
shock, fire and 
mechanical hazards 
only in accordance with UL60601, 
CAN/CSA C22.2 no 601.1
According to IEC 60601-1, CE models 
marked according to the Medical 
Device Directive 93/42/EEC (Device 
classification: II b),
Mobile sensor unit that is IEEE 
802.11b  (WLAN) compliant
Patient connection (sensor):
BF-type
Sensor technology:
CMOS Active Pixel technology 
(CMOS APS)
IMAGE QUALITY
Resolution:
26.3 lp/mm (theoretical)
Grayscale:
4096 gray levels.
Holders:
A full set of color coded KerrHawe and 
Sigma M sensor holders, with easy 
instructions, for comfortable and 
accurate positioning.
SOFTWARE MAIN 
FEATURES
Operating system:
Windows 2000 or XP (SP2)
Operating software:
CliniView version 6.4 or later
Full Mouth Series:
FMS, user defined
File sizes:
3,17 MB, uncompressed file,Sensor 
size1 4,96 MB, uncompressed 
file,Sensor size2
X-ray shield in the sensor:
Lead (Pb)
6.2
D
IMENSIONS
Outer dimensions:
Sensor unit:
102 x 60 x 22 mm
Sensor size 1:
26 x 37 mm
Sensor size 2:
31 x 44 mm
Active image area:
Sensor size 1:
30.02 x 19,95 mm
2
Sensor size 2:
36,48 x 25,84 mm
2
Approved
Reviewed: Anttila Mika Johannes 02.05.07 09:54:49
Approved: Levälampi Juhani Eemeli 02.05.07 12:02:32
See the PaloDEx Group Oy PDM system to determine the status of this document. Printed out: 04.05.07 12:38:31