Over
≤
Reason
Weight is exceed maximum capacity of the scale.
Trouble shooting
Do not exceed the weight limit.
If the load cell is damaged, the load cell should be replaced.
(2) CAL MODE
Err 21
≤
Reason
The resolution is set to be exceeded the limit 1/30,000.
Trouble shooting
Lower the resolution.
The resolution = Maximum weight/one division
Modify the maximum weight in CAL1 or modify the division in CAL2 so that the resolution
Modify the maximum weight in CAL1 or modify the division in CAL2 so that the resolution
should be below 1/30,000.
Err 22
≤
Reason
The weight for span calibration is set to be lower than 10 % of the maximum capacity
of the scale.
Trouble shooting
Set the weight for span calibration in CAL3 to be more than 10% of the
Set the weight for span calibration in CAL3 to be more than 10% of the
maximum capacity.
Err 23
≤
Reason
The weight for span calibration is set to be exceeded 100 % of the maximum capacity of the scale.
Trouble shooting
Set the weight for span calibration to be within the maximum capacity of the scale in CAL 1.
Trouble shooting
Set the weight for span calibration to be within the maximum capacity of the scale in CAL 1.
Notice : Specifications are subject to change for improvement