Electromagnetic Compatibility
Like other electrical medical equipment, the 
AMM215MWTD
 requires special precautions 
to ensure electromagnetic compatibility with other electrical medical devices. To ensure 
electromagnetic compatibility (EMC), the AMM215MWTD must be installed and operated 
according to the EMC information provided in this manual. 
Note 
 The AMM215MWTD has been designed and tested to comply with IEC 
60601-1-2:2001 requirements for EMC with other devices. 
Caution   Portable and mobile RF communications equipment may affect 
the normal function of the AMM215MWTD. 
Warning   Do not use cables or accessories other than those provided 
with the AMM215MWTD, as this may result in increased electromagnetic emis-
sions or decreased immunity to such emissions. 
Warning    If the AMM215MWTD is used adjacent to or stacked with 
other equipment, observe and verify normal operation of the AMM215MWTD 
in the configuration in which it will be used prior to using it in a surgical pro-
cedure. Consult the tables below for guidance in placing the AMM215MWTD. 
The Model AMM215MWTD is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM215MWTD should assure that it is used in such an environment. 
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The Model AMM215MWTD uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are not 
likely to cause any interference in nearby electronic equipment
RF emissions CISPR 11
Class B
AMM215MWTD is suitable for use in all establishments, 
including domestic establishments and those directly connected 
to the public low-voltage power supply network that supplies 
buildings used for domestic purposes
Harmonic emissions 
IEC61000-3-2
Class D
Voltage Fluctuations 
IEC61000-3-3
Complies
Manufacturer’s declaration - electromagnetic emission
The Model AMM215MWTD is intended for use in the electromagnetic environment specified below. 
The customer or the user of AMM215MWTD should assure that it is used in such an environment. 
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic 
environment - guidance
Electrostatic discharge 
(ESD) 
IEC61000-4-2
6 kV contact 
8 kV air 
6 kV contact 
8 kV air 
Floors should be wood, 
concrete or ceramic tile. 
If floors are covered with 
synthetic material, the relative 
humidity should be at least 
30%.
Electrical fast transient 
/burst IEC61000-4-4
2 kV for power supply lines
1 kV for input / output lines
2 kV for power supply lines
1 kV for input / output lines 
Mains power quality 
should be that of a typical 
commercial or hospital 
environment.
Surge 
IEC61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality 
should be that of a typical 
commercial or hospital 
environment.
Manufacturer’s declaration - electromagnetic immunity