Power frequency 
(50/60Hz) mag-
netic field 
IEC 61000-4-8 
3.0 A/m
3.0 A/m
Power frequency mag-
netic fields should be 
at levels characteristic 
of a typical location in 
a typical commercial or 
hospital environment.
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines IEC61000-4-11
<5% Ut (>95% dip in Ut) 
for 0.5 cycle
40% Ut (60% dip in Ut) 
for 5 cycles
70% Ut (30% dip in Ut) 
for 25 cycles
<5% Ut (>95% dip in Ut) 
for 5 sec.
<5% Ut (>95% dip in Ut) 
for 0.5 cycle
40% Ut (60% dip in Ut) 
for 5 cycles
70% Ut (30% dip in Ut) 
for 25 cycles
<5% Ut (>95% dip in Ut) 
for 5 sec.
Mains power quaility 
should be that of a typical 
commercial or hospital 
environment.  If the user 
of the AMM215MWTD 
image intensifier requires 
continued operation dur-
ing power mains interrup-
tions, it is recommended 
that the AMM215MWTD 
image intensifier be 
powered from an uninter-
ruptible power supply or 
a battery.
Note: Ut is the A.C. mains voltage prior to application of the test level.
The Model AMM215MWTD is intended for use in the electromagnetic environment specified below. 
The customer or the user of AMM215MWTD should ensure that it is used in such an environment. 
Immunity Test
IEC 60601 Test 
Level
Compliance Level
Electromagnetic environment 
- guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 
150 KHz to 80 MHz
3 V/m 
80MHz to 2.5 GHz
3 Vrms 
150 KHz to 80 MHz
3 V/m 
80MHz to 2.5 GHz
Portable and mobile RF communications 
equipment should be used no closer 
to any part of the 
AMM215MWTD
including cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of the 
transmitter.
Recommended Separation Distance
0.15 MHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where 
P is the maximum oputput power 
rating of the transmitter in watts (
W
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (
m).
Field strengths from fixed RF transmitters, 
as deter-mined by an electromagnetic 
site survey 
(a), should be less than the 
compliance level in each frequency range 
(b).Interferency may occur in the vicinity 
of equipment marked with the follwing 
symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and
             reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast, connot be predicted theoretically with accuracy.  
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered.  If the measured field strength in the location in which the AMM215MWTD is used exceeds the applicable RF 
compliance level above, the AMM215MWTD should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the AMM215MWTD.
(b) Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
Manufacturer’s declaration - electromagnetic immunity
AMM215MWTD   
29