Do not coil excess extension in front of the neurostimulator. Wrap excess
extension or leads around the perimeter (Figure 3) or behind the
neurostimulator to help minimize potential damage during neurostimulator
replacement surgery, help minimize potential kinking of the extension or lead,
and minimize interference with telemetry and recharge operation.
Figure 3. Wrap excess extension around the perimeter of the neurostimulator.
2. Use the suture holes in the connector block to secure the neurostimulator to the muscle
fascia with nonabsorbable silk.
Note: Secure the neurostimulator in the pocket to minimize movement or migration of
the neurostimulator.
Checking system integrity
The connections of the extensions and leads to the neurostimulator can be checked using the
clinician programmer. Refer to the programming guide for detailed programming
instructions.
w
Warning: To use the nonsterile programmer system components in a sterile field, place
a sterile barrier between the patient and system components to prevent infection. Do not
sterilize any components of the programmer system. Sterilization may damage the
components.
1. To ensure proper connection of each extension or lead to the neurostimulator, use the
clinician programmer to perform the lead insertion check.
2. If the lead insertion check results are not acceptable, refer to "Connecting the extension
Completing the implant procedure
1. Close and dress all incisions.
2. Turn off stimulation before sending your patient home.
3. Ensure that a patient control device is given to the patient.
12 English  97715 2012-06