    Fusion Wireless Recorder 
 
Setup Steps 
DO NOT ENROLL PATIENT IN SOFTWARE UNTIL INSTRUCED 
TO DO SO 
 
This manual is designed to allow a technician to follow the instructions 
page by page to setup the Fusion recorder.  Here is the general layout: 
 
The Fusion recorder is preprogrammed from the factory for default 
settings.  The device is fully programmable through the Fusion Wireless 
Monitoring System Software.  Please refer to the Fusion Wireless 
Monitoring System Software for programming capabilities and options 
 
1.  If the recorder contains batteries, and/or a cable remove them and 
insure the screen is blank before proceeding 
2.  Install fresh batteries into recorder.  Install only AA Lithium battery 
packs provided by Braemar.  Observe proper battery polarity when 
installing.  The battery indicator resets to indicate a full capacity 
battery each time the batteries are inserted.  For an accurate 
indication you must install fresh unused batteries. 
3.  Inspect LCD Screen. There are several screens that you may be 
greeted with when powering on a recorder.  This section covers each 
of those screens. 
 
“No cable” screen:  The recorder does not have any previous 
patient data and you can proceed to step 4. 
 
“Recording…” screen:  The recorder is still recording data, you 
must stop the recording and erase all the data before setting up 
the recorder for the next patient.     
 
Press the left and right buttons together and it will 
prompt you if you want to stop the recording.   
 
Press the Enter button to stop recording the data and 
the “Recording Complete” screen will appear. 
 
Continue to erase the data by following the “Recording 
Complete” screen steps. 
 
 
 
 
      
    Fusion Wireless Recorder 
 
 
“Recording Complete” screen:  The recorder still has data in it, 
you must erase all the data before setting up for the next patient.     
 
Press the left and right buttons together and it will 
prompt you to “Erase Data”.   
 
Press the Enter button to erase the data.  The display 
will have the message “Recorder Empty”. 
Continue setting up the recorder by following to the 
“Recorder Empty” screen steps. 
 
“Recorder Empty” screen:  The recorder is ready to be reset. 
Remove the batteries 
Return to step 2 
 
4.  Apply electrodes to patient 
5.  Connect snaps to electrodes 
6.  Plug cable into recorder 
7.  Observe LCD main menu screen 
8.  Verify lead status to be OK as described below 
9.  Verify ECG signal is OK as described below 
10.  Remove Batteries 
11.  Enroll patient on software via Enrollment Tab using the Fusion 
Wireless Monitoring System Software 
12.  Verify Enrollment by looking for yellow dot in the Patients Tab 
using the Fusion Wireless Monitoring System Software 
13.  Insert Batteries into the Fusion recorder 
14.  Verify the recording starts automatically by inspecting the LCD for 
the Recording…. screen 
15.  Verify yellow dot has been changed to two green dots in the Patients 
tab using the Fusion Wireless Monitoring System Software 
16.  Set up is complete and the recoding has been successfully started