Справочник Пользователя для Braemar Manufacturing LLC FUSIONMCT
Fusion Wireless Recorder
11
1. Patient cable connection
At this time, the patient leads should be connected to the electrodes,
the electrodes should be connected to the patient, and the patient
cable should be inserted into the recorder.
the electrodes should be connected to the patient, and the patient
cable should be inserted into the recorder.
Setup Menu Overview
2. Checking lead connections:
Using the left and right arrow buttons, select Lead Status
(
(
or
icons) and press the Enter button. Check the status of the
lead so make sure each of them has a good connection.
• Leads with good connections display
• Leads with poor connections display
• “Lead Error” will also display at the bottom of the screen
At least two leads must be connected to see the Lead Status. If there
are less than two leads present, a “No Cable” message is displayed.
Press the Enter button to return to the Setup Menu.
are less than two leads present, a “No Cable” message is displayed.
Press the Enter button to return to the Setup Menu.
Fusion Wireless Recorder
12
3. Checking ECG signals:
Using the left and right arrow buttons, select View ECG (
icon)
and press the Enter button.
You can change to another channel by pressing the left or right
buttons. If any channel is not available due to an error detecting a
lead, a “Lead Error” message will be displayed for only that affected
channels. If the cable becomes disconnected, all channels will
display “No Cable”.
buttons. If any channel is not available due to an error detecting a
lead, a “Lead Error” message will be displayed for only that affected
channels. If the cable becomes disconnected, all channels will
display “No Cable”.
To enter or exit zoom mode, press the left and right buttons
simultaneously for one second. A magnifying glass (
simultaneously for one second. A magnifying glass (
) will
appear next to the channel number in the upper right corner.
Pressing the right button will increase the amplitude of the signal,
while pressing left will decrease the amplitude of the signal. Pacer
pulse marks are displayed with a paced signal below the trace to
indicate each detection of a pacer pulse.
Pressing the right button will increase the amplitude of the signal,
while pressing left will decrease the amplitude of the signal. Pacer
pulse marks are displayed with a paced signal below the trace to
indicate each detection of a pacer pulse.
• Note that the gain setting only changes the display and not
the gain of the ECG signal stored in the recorder.
.5x gain
1x gain (default)
2x gain
Grid background indicates the display gain.
Press the Enter button to return to the Setup Menu.
General notes:
A. All button presses should beep to provide feedback for the user.
B. The backlight for the display is on while accessing the menus,
after recording a manual event, or after receiving a text
message.
message.
C. The backlight will remain off during normal recording
D. Lead loss and Pacer detection is on all the time.
E. The number of channels a patient cable contains will determine
the number of channels the recorder will record.