F.  Although the device detection algorithms are very sophisticated, 
there is no guarantee that the device will catch all episodes of 
arrhythmia. 
 
Start the Recording and Check Connection to Server 
 
The Recording screen should now be displayed.  An icon in the 
upper right corner indicates the type of connection that is being used 
to transfer the information to the Fusion Server. 
 
 
 
(not transmitting) 
Cellular 
 
 
Bluetooth modem 
USB cable 
 
A majority of the time the recorder will not be transmitting an so 
there will only be the battery icon in the upper right corner.   
 
Call the service center and verify the SN of the recorder has 
contacted the Fusion Server.  It is now the job of the service center 
complete enrollment of the patient with the recorder.  
 
 
 
 
      
    Fusion Wireless Recorder 
 
14 
Patient Operating Instructions 
 
Establishing the Home Link   
 
In accordance with FDA directive, Fusion Wireless Monitoring with 
patient alarm conditions requires the establishment of a Home Link 
alternative to cellular data communications.  The Fusion recorder kit 
contains Bluetooth wireless hardware that must be connected to a normal 
RJ-11C telephone jack.  The jack used should typically be located on the 
nightstand or near where the patient will spend most of their time during 
the procedure.  Patients inside the home location should be in Home Link 
wireless contact if they are within approximately 75 feet of the system.   
Note: This modem has an actual line of site range of 328 feet but due to 
walls and other structural impedance the modem should be placed within 
100 feet.   
 
The Fusion recorder and Home Link modem connections are preset at 
the factory and do not need any user configuration.