Справочник Пользователя для Agfa Gevaert N V 5364A

Скачать
Страница из 266
195
2900H EN 20071108
Troubleshooting
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
Checking the status indicator LED
The LED on the front panel indicates the Drystar 5500 operating status. Use
the following table to determine the status of your printer.
Color
Light
Status
Receive 
jobs?
Printing 
OK?
Action
Green
Constant
Ready (standby)
Yes
Yes
Proceed.
Blinking
Start-up 
sequence
No
No
Wait.
Receiving jobs
Yes
Yes
Calculating
Yes
Yes
Printing
Yes
Yes
Calibrating
Yes
No
Installation
No
No
Proceed with 
the activity.
Key-operator
Yes/No
Yes/No
Red
Blinking
Service mode
No
No
Upper/Lower 
Input tray empty
Yes
Yes/No
Refer to 
on page
Film jam
Yes
No
Refer to 
page
Warning
Yes
Yes
Refer to 
page
Constant
Error status
No
No
Refer to 
page