Справочник Пользователя для Agfa Gevaert N V 5364A

Скачать
Страница из 266
254
2900H EN 20071108
Remarks for HF-emission and immunity
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
Remarks for HF-emission and immunity
This device is intended for operation in the electromagnetic 
environment given below. The user of the device should ensure 
that it is used in such an environment.
This device was tested for a normal hospital environment as described above.
Nevertheless the HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
Transmission Measurements
Agreement
Electromagnetic Environment 
Guidelines
High frequency transmissions in 
accordance with CISPR 11
Group 1
The device uses high frequency energy 
exclusively for its internal functions. For this 
reason, its high frequency transmission is 
very low and it is improbable that 
neighboring electronic equipment will be 
disrupted. 
High frequency transmissions in 
accordance with CISPR 11
Class A
The device is intended for use in an 
environment that do not include living areas 
and in areas directly connected to a public 
supply network that also supplies buildings 
that are used for domestic purposes
Excess oscillations in accordance 
with IEC 61000-3-2
Class A
Voltage fluctuations / flickering in 
accordance with IEC 61000-3-3
Fulfilled