Remarks for HF-emission and immunity
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
This device is intended for operation in the electromagnetic 
environment given below. The user of the device should ensure 
that it is used in such an environment.
 Resistance to 
Jamming Test
IEC 60601 
Test Level
Level of 
Agreement
Electromagnetic 
Environment Guidelines
Discharge of 
static electricity 
in accordance 
with IEC 61000-
4-2
+ 4 kV contact 
discharge
+ 8 kV air 
discharge
+ 6 kV contact 
discharge
+ 8 kV air 
discharge
Floors should consist of 
wood, concrete or ceramic 
tiles. The relative humidity 
must be at least 30%, if the 
floor is made of synthetic 
material.
Fast transient 
electrical 
disturbance 
variables / bursts 
in accordance 
with IEC 61000-
4-4
+ 1 kV mains
+ 0,5 kV data 
lines
+ 1 kV mains
+ 0,5 kV data 
lines
The quality of the voltage 
supplied should correspond 
to a typical commercial or 
clinical environment.
Impulse voltages 
(surges) in 
accordance with 
IEC 61000-4-5
+ 1 kV push-pull 
voltage
+ 2 kV common 
mode voltage
+ 1 kV push-pull 
voltage
+ 2 kV common 
mode voltage
The quality of the voltage 
supplied should correspond to 
that of a typical commercial or 
clinical environment.
Voltage 
breakthroughs, 
short term 
interruptions and 
variations in the 
voltage supplied in 
accordance with 
IEC 61000-4-11
• <  5%  U
(> 95% 
breakthrough of 
U
r
) for ½ period
• 40%  U
(> 60% 
breakthrough of 
U
r
) for 5 periods
• 70%  U
(30% 
breakthrough of 
U
r
) for 25 periods
• <  5%  U
(95% 
breakthrough of 
U
r
) for 5 s
• <  5%  U
(> 95% 
breakthrough of 
U
r
) for ½ period
• 40%  U
(> 60% 
breakthrough of 
U
r
) for 5 periods
• 70%  U
(30% 
breakthrough of 
U
r
) for 25 periods
• <  5%  U
(95% 
breakthrough of 
U
r
) for 5 s
The quality of the voltage supply 
should correspond to that of a 
typical commercial or clinical 
environment.
If the user wants the device to 
work continuously, even when 
the energy supply is interrupted, 
it is recommended to use an 
energy supply free of 
interruptions or a battery.
Magnetic field at 
the supply 
frequency (50/60 
Hz) in accordance 
with IEC 61000-4-8
3 A/m
10 A/m
Magnetic field at the network 
frequency should correspond to 
the typical values as they are in a 
commercial and clinical 
environment.
• REMARK  :  U
r
 is the alternating current in the network
 
before the application of the test level.