Short Range Programmer VRT—User Guide   
 
 
P/N 52806T100 Rev 1 April 2003  
  AMCO Automated Systems
 
 
Using Read Pre-divider (and CF) Mode 
 
This mode reads the current sub-counts and pre-divider along with the 
current compensation factor (V9 and VRT). 
 
NOTE: The CF displays with the pre-divider only if the transponder 
firmware supports CF and this can only be determined if the transponder 
is interrogated first in the Find mode. 
 
1.  Press the MODE button repeatedly until the display shows either of 
the two “Read Pre-divider” screens below: 
 
•  For transponders accessed directly through Pre-divider mode: 
 
•  For V9 and VRT transponders interrogated in Find mode: 
 
 
2.  If the serial number represents the target transponder, go to step 4. 
 
3.   If Find mode was not used, or the displayed serial number is not the 
serial number of the target transponder, enter the serial number of 
the desired transponder. 
 
4.  Press either of the two INT (interrogate) buttons to read the pre-
divider (and CF if applicable). 
 
•  If interrogation is successful, the screen will display the serial 
number and pre-divider (and CF if applicable) as shown in the 
examples below: 
 
•  The pre-divider is displayed in the form SSS/PPP where SSS is 
the current sub-count and PPP is the actual pre-divider. 
 
 
READ PRE-DIVIDER 
 Serial No..  03571641 
PRE-DIV...  =  001/100 
READ PRE-DIVIDER 
 Serial No..  03571641 
 
READ PRE-DIV and CF 
 Serial No..  03571641