15. Specification 
84 
 
Time Constant 
0.4 sec ±25% 
Switching and available to set 0.1 sec ±25% 
Pacemaker Pulse 
Detection/ Rejection: 
Comply with ANSI/AAMI EC13 Pacemaker pulse 
rejection capability 
Protection to 
Defibrillation: 
Meet the requirement of IEC60601-2-27 
 
Respiration (Impedance Method) 
Accuracy of Sensitivity: 
10mm/1Ω ±2mm   
(When standard Impedance is 480Ω.) 
Display Coverage of Respiration Rate:0, 9 to 150Bpm 
Display Error of Respiration Rate: 
3Bpm 
Measured Current of Respiration: 
Below 100μA (at 42kHz) 
 
SpO
2
 
LX-7230N 
SpO
2
 Measurement Range: 
1 to 100%   
Resolution: 1% 
Measurement Accuracy: 
Accuracy of measurement with SpO
2
 
Probe is as follows. 
 
SpO
2
 Probe 
Measurement 
Accuracy (±1SD)
OxiMax
® 
 MAX-I 
±2% 
OxiMax
®  
MAX-P ±2% 
OxiMax
®  
MAX-A/AL ±2% 
Durasensor
®  
DS-100A ±3% 
OxiMax
®  
MAX-R ±3.5% 
OxiMax
®  
MAX-FAST ±2% 
OxiMax
®  
MAX-N ±2% 
(When SpO
2
 is 70 to 100%. Less than 70% is not specified.) 
 
 
Wavelength: 
Approx. 660nm (Red light) 
Approx. 890nm (Infrared light)