Справочник Пользователя для Fukuda Denshi Co Ltd LX7230KM
15. Specification
84
Time Constant
0.4 sec ±25%
Switching and available to set 0.1 sec ±25%
Switching and available to set 0.1 sec ±25%
Pacemaker Pulse
Detection/ Rejection:
Detection/ Rejection:
Comply with ANSI/AAMI EC13 Pacemaker pulse
rejection capability
rejection capability
Protection to
Defibrillation:
Defibrillation:
Meet the requirement of IEC60601-2-27
Respiration (Impedance Method)
Accuracy of Sensitivity:
10mm/1Ω ±2mm
(When standard Impedance is 480Ω.)
(When standard Impedance is 480Ω.)
Display Coverage of Respiration Rate:0, 9 to 150Bpm
Display Error of Respiration Rate:
Display Error of Respiration Rate:
3Bpm
Measured Current of Respiration:
Below 100μA (at 42kHz)
SpO
2
LX-7230N
SpO
2
Measurement Range:
1 to 100%
Resolution: 1%
Measurement Accuracy:
Measurement Accuracy:
Accuracy of measurement with SpO
2
Probe is as follows.
SpO
2
Probe
Measurement
Accuracy (±1SD)
OxiMax
®
MAX-I
±2%
OxiMax
®
MAX-P ±2%
OxiMax
®
MAX-A/AL ±2%
Durasensor
®
DS-100A ±3%
OxiMax
®
MAX-R ±3.5%
OxiMax
®
MAX-FAST ±2%
OxiMax
®
MAX-N ±2%
(When SpO
2
is 70 to 100%. Less than 70% is not specified.)
Wavelength:
Approx. 660nm (Red light)
Approx. 890nm (Infrared light)
Approx. 890nm (Infrared light)