15. Specification
86
Transmission Method
Modulation Mode:
Digital, Frequency shift keying
Frequency:
608 to 614MHz
Oscillation Method:
PLL Synthesizer method by crystal control
Channel Spacing:
12.5kHz
Occupied Frequency
Bandwidth:
Bandwidth:
8.5kHz
RF output power:
1mW ±2dB
Transmitting Antenna:
ECG lead cable and/or SpO
2
Probe
Safety
General Standard: IEC 60601-1:1988 +A1: 1991 +A2: 1995
(Medical electrical equipment – Part 1: General
requirements for safety)
requirements for safety)
EMC Standard:
IEC 60601-1-2: 2007
(Medical electrical equipment – Part 1: General
requirements for safety – 2. Collateral standard:
Electromagnetic compatibility – Requirements and tests)
(Medical electrical equipment – Part 1: General
requirements for safety – 2. Collateral standard:
Electromagnetic compatibility – Requirements and tests)
The class of
protection
against electric
shock
protection
against electric
shock
Internally Powered Equipment
The type of
protection
against electric
shock
protection
against electric
shock
Type CF Applied part
Operating Environment
Temperature:
10 to 40
C
Humidity:
30 to 85% RH (No condensation)
Vibration/Shock:
Comply with IEC60068-2-64:1987, IEC60068-2-32:1975,
IEC60068-2-6:1995
IEC60068-2-6:1995
Transport / Storage Environment
Temperature:
-10 to 60
C
Humidity:
10 to 95% RH (No condensation)