6. SpO
2
 Monitor (LX-7230KM) 
39 
 
  Applying Spot Check Probe (SP-5C)   
 
The SP-5C is suitable for short-term measurement, and should be clipped to a 
fingertip of an adult/ pediatric who is 30kg and above. 
It should be positioned so that the LED (light source) is at the base of the 
fingernail. 
 
SP-5C 
Light Source
Sensor
Proper attachment
Finger is not sufficiently
inserted.
Finger is inserted too far.
Measurement cannot be taken
  
 
WARNING 
  Precautions for Use of Spot Check Probe (SP-5C) 
  The SP-5C is designed for short-term measurement. If measurements are taken 
continuously over a long period of time, change the measuring finger every 2 
hours (recommended wearing hours) to prevent low-temperature burn. Especially 
for continuous use on patient with peripheral circulatory disturbance, change the 
measuring finger more frequently. 
  Do not use tape to fix the probe. It may cause edema or congestion. 
 
CAUTION 
  Precautions for Use of Spot Check Probe (SP-5C) 
  The probe should be positioned so that the light source (LED) is at the base of 
the fingernail. Take care not to insert a finger too deep to prevent injury. 
  When removing the probe, do not pull on the cord. 
  Cover the probe with a black cover if it is exposed to high-intensity light such as 
direct light and surgical light to prevent measurement error. 
  The SP-5C is designed for use on finger of an adult/ pediatric who is 30kg and 
above. Do not use on other parts. 
  Before attaching the probe to a patient, clean the probe using a cloth moistened 
with sterilizing alcohol. 
  Remove nail polish before taking measurements to enable stable 
measurements. 
  It may not be possible to take measurements in case of an excessively high 
hematocrit value. In such case, reattach the probe to a thinner finger. 
  Take note that it may not be possible to take measurements in place with 
vibrations or while walking. 
 
After attaching the probe, make sure that there is no failure message and the 
pulse wave level on the LCD is upper than 2. If not upper than level 2, it may 
measure incorrectly. In such case, warm the finger to improve blood circulation.