6. SpO
2
 Monitor (LX-7230KM) 
40 
 
  Applying Universal Probe (UD-5C) 
 
The UD-5C is suitable for long-term measurement, and can be attached to a 
fingertip of an adult/ pediatric who is 40kg and above. 
 
1. 
Fix the probe with tape. Before attaching the probe to a patient, clean the 
probe using a cloth moistened with sterilizing alcohol. 
2. 
Make sure that the two pads are aligned. 
3. 
Attach the pad with the LED to the base of the fingernail and the sensor 
pad to the opposite side of the finger. 
 
Light source (LED)
UD-5C LED 
mark 
             
                         
 
Sensor 
 
WARNING 
  Precautions for Use of Universal Probe (UD-5C) 
  Although UD-5C is suitable for long-term measurement, change the measuring 
site every 8 hours (recommended wearing hours) to prevent rash or a low-
temperature burn. Be especially careful for continuous use on a patient with 
peripheral circulatory disturbance by changing the measuring site frequently. 
  If the use of adhesive tape causes skin irritation, it should be discontinued. 
  When securing the probe with bandage tape, do not tighten it too hard as this 
may cause edema or congestion. 
 
When removing the tape, be careful not to damage the skin.
 
 
CAUTION 
  Precautions for Use of Universal Probe (UD-5C) 
  When removing the probe, do not pull on the cord, and remove the tape. Do not 
apply excessive force to remove the tape. It may break the probe. 
  Cover the probe with a black cover if it is exposed to high-intensity light such as 
direct light and surgical light to prevent measurement error. 
  Remove nail polish before taking measurements to enable stable 
measurements. 
  It may not be possible to take measurements in case of an excessively high 
hematocrit value. In such case, reattach the probe to a thinner finger. 
 
After attaching the probe, make sure that there is no failure message and the 
pulse wave level on the LCD is upper than 2. If not upper than level 2, it may 
measure incorrectly. In such case, warm the finger to improve blood circulation.