Справочник Пользователя для Fukuda Denshi Co Ltd LX7230KM
8. Operation
61
●
SpO
2
Display Screen
SpO
2
display screen depends on the transmitter LX-7230N or LX-7230KM.
LX-7230N:
LX-7230N has no setup item in the SpO
2
display screen.
LX-7230KM:
LX-7230KM has a setup item to select the SpO
2
measurement mode (continuous or intermittent) in the SpO
2
display screen.
【
Setting Method】
How to enter the setup mode:
Press and hold the Enter button for 2 seconds in the SpO
Press and hold the Enter button for 2 seconds in the SpO
2
display screen.
<<Changing SpO
2
Measurement Mode>>
Indicates the current measurement
mode. (In the SpO
2
continuous
measurement mode, nothing is
indicated. In the SpO
indicated. In the SpO
2
intermittent
measurement mode, “Int” is
indicated.)
Pressing the Enter button will
sequentially change modes.
indicated.)
Pressing the Enter button will
sequentially change modes.
Mode Continuous intermittent
Continuous
<< Returning to SpO
2
display screen>>
Press the ▽ button to highlight
the Return button.
Press the Enter button to return to
the SpO
the Return button.
Press the Enter button to return to
the SpO
2
display screen.
CAUTION
In the SpO
2
intermittent mode, the SpO
2
is measured at 25 seconds
interval. Even when the SpO
2
value suddenly changes during this
interval, the LX-7230 and receiving monitor will only show the last SpO
2
measurement value. For a patient with the possibility of sudden change,
select SpO
select SpO
2
continuous mode.
Pulse wave and pulse wave level meter are not displayed on the LCD
during the SpO
2
intermittent measurement mode. On the screen of the
receiving monitor, pulse wave is not displayed as well.
Make sure that SpO
2
is measured correctly, and the SpO
2
value and
pulse rate are updated at 25 seconds interval on the LCD or on the
screen of the receiving monitor.
screen of the receiving monitor.