8. Operation 
61 
 
SpO
2
 Display Screen 
SpO
2
 display screen depends on the transmitter LX-7230N or LX-7230KM. 
 
LX-7230N: 
LX-7230N has no setup item in the SpO
2
 display screen. 
LX-7230KM: 
LX-7230KM has a setup item to select the SpO
measurement mode (continuous or intermittent) in the SpO
2
 
display screen. 
 
Setting Method】 
How to enter the setup mode: 
Press and hold the Enter button for 2 seconds in the SpO
2
 display screen. 
 
<<Changing SpO
Measurement Mode>> 
 
Indicates the current measurement 
mode. (In the SpO
continuous 
measurement mode, nothing is 
indicated. In the SpO
intermittent 
measurement mode, “Int” is 
indicated.)  
 
Pressing the Enter button will 
sequentially change modes. 
Mode Continuous  intermittent 
 Continuous 
 
<< Returning to SpO
2
 display screen>> 
 
Press the  ▽  button to highlight 
the Return button. 
Press the Enter button to return to 
the SpO
2
 display screen. 
 
CAUTION 
  In the SpO
intermittent mode, the SpO
2
 is measured at 25 seconds 
interval. Even when the SpO
2
 value suddenly changes during this 
interval, the LX-7230 and receiving monitor will only show the last SpO
2
 
measurement value. For a patient with the possibility of sudden change, 
select SpO
continuous mode. 
  Pulse wave and pulse wave level meter are not displayed on the LCD 
during the SpO
intermittent measurement mode. On the screen of the 
receiving monitor, pulse wave is not displayed as well. 
  Make sure that SpO
is measured correctly, and the SpO
value and 
pulse rate are updated at 25 seconds interval on the LCD or on the 
screen of the receiving monitor.