9. Other Setting Items 
The following settings are available for the LX-7230 depending on the use and 
condition of the patient. For details of the settings, contact our service 
representative. 
 
Items Selection 
Default
Backup 
Time Constant 
0.4 sec., 0.1 sec. 
0.4 sec.
Yes 
Detection Sensitivity of 
Pacemaker Pulse 
Low, Mid, High 
Mid 
Yes 
Respiration Detection 
Signal 
ON, OFF 
ON 
Yes 
LCD Contrast 
8 steps 
Yes 
Transmitter Channel 
One from the 
following channels. 
 
0801 to 0879 
0900 to 0979 
1000 to 1079 
1100 to 1179 
1200 to 1279 
1300 to 1379 
 
1100 Yes 
Group ID 
One from 00 to 63 
00 
Yes 
 
  Changing the Time Constant 
The default setting of the time constant is “0.4 seconds”. 
If a stable monitoring is difficult with excessive change in the baseline due to 
excessive body motion of the patient or an interference noise, such as AC 
frequency, by changing the time constant to “0.1 second” the monitoring may 
become relatively stable. 
 
For details of the setting change, contact your local Fukuda Denshi service 
representative. 
 
CAUTION 
  When changing the time constant to “0.1 seconds”, the lower frequency 
characteristic becomes 1.6Hz ±25%. This setup does not meet IEC 
60601-2-27 standard. It may lead to a change in the ECG waveform and 
ST measurement value may be especially affected. Fukuda Denshi 
recommends “0.4 seconds” setting in normal use. 
  The LCD screen in normal use does not indicate the selection of time 
constant. Make sure to take measures, such as marking on the LX-
7230, to distinguish whether the selection of time constant is changed.