Справочник Пользователя для Fukuda Denshi Co Ltd LX7230KM
9. Other Setting Items
65
■ Changing the Respiration Detection Signal ON/OFF
The default setting of the respiration detection signal is “ON”.
The respiration waveform can be detected when the setting of the respiration
detection signal is turned “ON”.
The respiration waveform can be detected when the setting of the respiration
detection signal is turned “ON”.
WARNING
If the LX-7230 is used with minute ventilation rate-adaptive implantable
pacemaker, the respiration detection signal may cause the pacemaker to
pace at its maximum programmed rate. If such event occurs, change the
setting to “OFF” to prevent an occurrence of erroneous pacing rate.
pacemaker, the respiration detection signal may cause the pacemaker to
pace at its maximum programmed rate. If such event occurs, change the
setting to “OFF” to prevent an occurrence of erroneous pacing rate.
For details of the setting change, contact your local Fukuda Denshi service
representative.
CAUTION
The respiration waveform cannot be measured if the setting of the
respiration detection signal is turned “OFF”.
Make sure to turn OFF the respiration measurement function on the
receiving monitor to prevent an erroneous detection of the respiration
alarm (on the receiving monitor side).
alarm (on the receiving monitor side).
The LCD screen in normal use does not indicate the setting status of the
respiration detection signal ON/OFF. Make sure to take measures, such
as marking on the LX-7230, to distinguish whether the setting of the
respiration detection signal ON/OFF is changed.
as marking on the LX-7230, to distinguish whether the setting of the
respiration detection signal ON/OFF is changed.
■ Changing the LCD Contrast
The LCD display contrast of the LX-7230 can be changed in 8 steps.
For details of the setting change, contact your local Fukuda Denshi service
representative.
For details of the setting change, contact your local Fukuda Denshi service
representative.