9. Other Setting Items 
65 
 
  Changing the Respiration Detection Signal ON/OFF 
The default setting of the respiration detection signal is “ON”. 
The respiration waveform can be detected when the setting of the respiration 
detection signal is turned “ON”. 
 
WARNING 
If the LX-7230 is used with minute ventilation rate-adaptive implantable 
pacemaker, the respiration detection signal may cause the pacemaker to 
pace at its maximum programmed rate. If such event occurs, change the 
setting to “OFF” to prevent an occurrence of erroneous pacing rate. 
 
For details of the setting change, contact your local Fukuda Denshi service 
representative. 
 
CAUTION 
  The respiration waveform cannot be measured if the setting of the 
respiration detection signal is turned “OFF”. 
  Make sure to turn OFF the respiration measurement function on the 
receiving monitor to prevent an erroneous detection of the respiration 
alarm (on the receiving monitor side). 
  The LCD screen in normal use does not indicate the setting status of the 
respiration detection signal ON/OFF. Make sure to take measures, such 
as marking on the LX-7230, to distinguish whether the setting of the 
respiration detection signal ON/OFF is changed. 
 
  Changing the LCD Contrast 
The LCD display contrast of the LX-7230 can be changed in 8 steps. 
 
For details of the setting change, contact your local Fukuda Denshi service 
representative.