Operator's Manual  ZM-940PA
The cuff is inflated and the inflation pressure is displayed on the screen.
In manual mode: Measurement is performed once.
In STAT mode:
Measurement is performed repeatedly for 15 minutes.
In auto mode:
The first measurement is performed when the NIBP START/STOP key
is pressed.  The second measurement is performed when the current time
in the transmitter reaches the selected time interval.
To stop measurement during measurement, press the NIBP START/STOP key again.
In STAT mode, after completing the STAT measurement, the measurement mode changes to
the mode set for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen.
In auto mode, to stop measurement in auto mode, change the mode to manual.  To cancel
one measurement, press the NIBP START/STOP key during measurement.
After the measurement is complete, the measured data is displayed on the screen and is
transmitted to the monitor.
Inflation pressure
When SpO
2
 is not monitored, the pulse rate at the end of NIBP measurement is displayed.
During auto mode measurement, the measurement mode can be changed.  During the interval,
press the NIBP INTERVAL key to change the mode.  When “MANUAL” is displayed for more
than one second, the measurement in auto mode is stopped.
A buzzer can be set to sound at the start and end of NIBP measurement.  Refer to the “Changing
Parameter Setup Settings” section.