Operator's Manual  ZM-940PA
61
Data Display After NIBP Measurement
When the time set at “OLD NIBP DATA” on the PARAMETER SETUP screen elapses after
the last measurement, the NIBP data is dimmed or hidden.  Whether to dim or hide the old data
can also be selected at “OLD NIBP DATA”.  Refer to the “Changing Parameter Setup Settings”
section.
Data Display on the Receiving Monitor
When the “Low battery 
” message is displayed on the receiving monitor, NIBP might not
have been measured according to the NIBP interval setting.  Therefore, the NIBP data displayed
on the receiving monitor might not be updated.  In this case, check the measurement time of the
NIBP data displayed on the receiving monitor.
Monitoring SpO
2
 during NIBP Measurement
When the SpO
2
 probe is attached to the same limb as the NIBP cuff, the blood flow decreases
during NIBP measurement and pulse wave cannot be detected and SpO
2
 cannot be monitored
properly.   When “INHIBIT SpO
2
 DURING NIBP” on the PARAMETER SETUP screen is set
to ON (factory default setting), SpO
2
 monitoring is paused during NIBP measurement to avoid
SpO
2
 alarm occurrence.  However, when monitoring SpO
2
 on the same limb as the NIBP, be
careful when reading SpO
2
 values.
ECG and Respiration Monitoring
When the electrodes are attached and the ECG leads are connected to the electrodes, heart rate,
ECG waveform, respiration rate and respiration waveform appear on the monitor.
When 6 leads are used on this transmitter, up to 8 lead (I, II, III, aVR, aVL, aVF, Va and Vb) of
ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured.
When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving
monitor. Refer to the operator’s manual of the monitor for details.
Respiration is monitored by measuring changes in impedance between the RA and LL ECG
electrodes. This transmitter sends the changes in impedance to the monitor as a respiration
waveform. The monitor displays the respiration waveform and calculates respiration rate. Refer
to the operator’s manual of the monitor for details.
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac
Monitoring and Diagnostic Equipment*
The bioelectric impedance measurement sensor of a minute ventilation rate-