Справочник Пользователя для Biolog-id SSTRBIOLOGID
Smart Storage for fridge : user manual
Backup date
12/05/2017 17:14:00
Version
0.10
Author
Loïc Marouteau
Manual validation date: 19/04/2017
27 sur 29
Red drawer space LED
Possible causes
Action
Data cannot be written
Move the bag to a different space and try again
The drawer has been open for longer than 4
minutes
minutes
Close the drawer
Communication with processor lost
Switch to maintenance mode, then restart the
processor
processor
Red processor LED
Possible causes
Action
CAN bus communication lost
Switch to maintenance mode and restart the
processor
processor
The CAN bus power supply has short-circuited
Switch to maintenance mode and restart the
processor
processor
The processor probe has become disconnected Connect the temperature probe
The Ethernet network has become
disconnected
The Ethernet network has become
disconnected
Unplug the Ethernet cable and plug it back in again
Insufficient SD card memory
Check that a notification has been sent to the third
party system. Contact the administrator.
party system. Contact the administrator.
Power supply fault on at least one drawer
Check that a notification has been sent to the third
party system. Contact the administrator.
party system. Contact the administrator.
Battery charger fault
Check that a notification has been sent to the third
party system. Contact the administrator.
party system. Contact the administrator.
When a red LED appears, try to detect the cause of the fault and eliminate it as quickly as
possible.
Contact the supplier of your device if it malfunctions.
Never carry out repairs or modifications without first obtaining authorisation from Biolog-Id.
Before carrying out any maintenance operations, the SST-R must be switched to
maintenance mode (cf HMI Manual).
possible.
Contact the supplier of your device if it malfunctions.
Never carry out repairs or modifications without first obtaining authorisation from Biolog-Id.
Before carrying out any maintenance operations, the SST-R must be switched to
maintenance mode (cf HMI Manual).
7 Warranty
Failure to observe any of the recommendations will void the warranty.