Справочник Пользователя для Guided Therapeutics Inc 13500
November 2013, CNDS-0-635-MAN_3
Page 43 of 78
8. After device is visibly clean, dry completely with a clean task wipe.
9. Place cleaned HHU on disposable task wipe.
HHU Window
1. Visually inspect the HHU Window for visible soil or streaks.
2. If the window is streaked, wipe the HHU window with disposable task wipes moistened with detergent solution.
3. Discard wipes as they become contaminated.
4. Visually inspect the HHU Window for soil.
5. Repeat wiping until window is not soiled.
6. Wipe window with pre-moistened 70% IPA wipe or 70% IPA cotton swab applicator.
7. Visually inspect the HHU Window for streaks. Repeat IPA wiping until the window is dry without streaks. Allow to air
dry at ambient temperature (68° F) /(20° C )
Umbilical
1. Place HHU on a clean disposable task wipe or disinfected cleaning container, such as a metal surgical tray.
2. Keep the HHU handle pointing toward the table/floor to avoid cleaning fluid penetrating the umbilical-handle
interface.
3. Wet a disposable task wipe with the detergent solution. Using disposable task wipes, remove all gross visible soil
from the umbilical cable. Discard each wipe as it becomes dirty. It is best to wipe and discard, using a new disposable
task wipe for each removal. Wipe away from the HHU, toward the handle.
task wipe for each removal. Wipe away from the HHU, toward the handle.
4. Once gross visible soil is removed, thoroughly spray detergent solution onto a disposable task wipe. Apply wipe
directly onto the areas to be cleaned. Allow device to be saturated with solution to dissolve any remaining organic
soil, loosen attachment of bioburden and thoroughly wet the device surface. Wet surface with soaked wipes at least
2 minutes to ensure that soil is hydrated for easier removal. If matter has dried on, continue to soak surface with
new wet wipes until visibly clean. Pay attention to the HHU umbilical cable interface.
soil, loosen attachment of bioburden and thoroughly wet the device surface. Wet surface with soaked wipes at least
2 minutes to ensure that soil is hydrated for easier removal. If matter has dried on, continue to soak surface with
new wet wipes until visibly clean. Pay attention to the HHU umbilical cable interface.
5. Using detergent solution and disposable task wipes or a pre-saturated detergent sponge, manually scrub the
umbilical cable at least 2 minutes. Dispose of wipe/sponges as needed. Rinse device by wiping with wipes moistened
with tap water at (65° F) /(18° C )Under normal office lighting, visually inspect the cable to ensure soil has been
removed. If visible soil remains, clean the cable again as directed above.
with tap water at (65° F) /(18° C )Under normal office lighting, visually inspect the cable to ensure soil has been
removed. If visible soil remains, clean the cable again as directed above.
6. After cable is visibly clean, dry completely with a clean, dry disposable task wipe
BU-Dock
1. Wet a disposable task wipe with the detergent solution. Using disposable task wipes, remove all gross visible soil
from the contaminated area (docking station). Discard each wipe as it becomes dirty. It is best to wipe and discard,
using a new disposable task wipe for each removal.
using a new disposable task wipe for each removal.
2. Once gross visible soil is removed, thoroughly spray detergent solution onto a disposable task wipe. Apply wipe
directly onto the areas to be cleaned. Allow device to be saturated with solution to dissolve any remaining organic
soil, loosen attachment of bioburden and thoroughly wet the device surface. Wet surface with soaked wipes at least
2 minutes to ensure that soil is hydrated for easier removal. If matter has dried on, continue to soak surface with
new wet wipes until visibly clean.
soil, loosen attachment of bioburden and thoroughly wet the device surface. Wet surface with soaked wipes at least
2 minutes to ensure that soil is hydrated for easier removal. If matter has dried on, continue to soak surface with
new wet wipes until visibly clean.
3. Using detergent solution and disposable task wipes or pre-soaked detergent sponge, manually scrub the device until
visible soil is removed— at least 2 minutes. Dispose of wipes/sponge as needed. Rinse device by wiping with wipes
moistened with warm tap water at (68° F -104° F) /(20° C – 40° C)
moistened with warm tap water at (68° F -104° F) /(20° C – 40° C)
4. Under normal office lighting, visually inspect the device to ensure soil has been removed. If visible soil remains, clean
the device again as directed above.
5. After device is visibly clean, dry completely with a clean task wipe.
Touchscreen
1. Caution: DO NOT spray detergent solution directly onto monitor. Rear of monitor is not protected against fluid
intrusion. Monitor should be cleaned by spraying detergent onto wipes, then using moist wipe to clean the surfaces.
2. Wet a disposable task wipe with the detergent solution. Using disposable task wipes, remove all gross visible soil
from the contaminated touchscreen. Discard each wipe as it becomes dirty. It is best to wipe and discard, using a
new disposable task wipe for each removal.
new disposable task wipe for each removal.
3. Once gross visible soil is removed, thoroughly spray detergent solution onto a disposable task wipe. Apply wipe
directly onto the areas to be cleaned. Allow device to be saturated with solution to dissolve any remaining organic
soil; loosen attachment of bioburden and thoroughly wet the device surface. Wet surface with soaked wipes at least
2 minutes to ensure that soil is hydrated for easier removal. If matter has dried on, continue to soak surface with
new wet wipes until visibly clean. Pay attention to monitor touchscreen–bezel interface.
soil; loosen attachment of bioburden and thoroughly wet the device surface. Wet surface with soaked wipes at least
2 minutes to ensure that soil is hydrated for easier removal. If matter has dried on, continue to soak surface with
new wet wipes until visibly clean. Pay attention to monitor touchscreen–bezel interface.
4. Using detergent solution and disposable task wipes, manually scrub the device until visible soil is removed, at least 2
minutes. Dispose of wipes as needed. Rinse device by wiping with wipes moistened with warm tap water at (68° F -
104° F) /(20° C – 40° C)Under normal office lighting, visually inspect the device to ensure soil has been removed. If
visible soil remains, clean the device again as directed above.
104° F) /(20° C – 40° C)Under normal office lighting, visually inspect the device to ensure soil has been removed. If
visible soil remains, clean the device again as directed above.
5. After device is visibly clean, dry completely with a clean task wipe.
BU-Touchscreen
bezel
bezel
1. Caution: DO NOT spray detergent solution directly onto monitor. Rear of monitor is not protected against fluid
intrusion. Monitor should be cleaned by spraying detergent onto wipes, then using moist wipe to clean the surfaces.
2. Wet a disposable task wipe with the detergent solution. Using disposable task wipes, remove all gross visible soil
from the contaminated monitor bezel. Discard each wipe as it becomes dirty. It is best to wipe and discard, using a