Intermediate Level Disinfectant (ILD) solution  
Quaternary ammonium compounds and isopropyl alcohol in aqueous solution 
 Example: CaviCide  
Nominally 0.25% n-Alkyl dimethyl benzyl ammonium chlorides  and 10-20% isopropyl alcohol in 
aqueous solution 
 
Intermediate Level Disinfectant Wipes 
Quaternary ammonium compounds and isopropyl alcohol in pre-saturated wipes 
 Example: CaviWipes, Sani-Wipes 
Nominally 0.25% n-Alkyl dimethyl benzyl ammonium chlorides  and 10-20% isopropyl alcohol in 
aqueous solution 
 
Pre-moistened 70% IPA wipes or 70% IPA cotton swab applicator. 
 
10.3.3 Disinfection Instructions  
 
CAUTION: The HHU and Base Unit are IP20. The units are not designed for submersion or 
excessive fluid ingress. Clean, rinse and disinfect via wiping with task wipes moistened with 
cleaning solution, rinse water or disinfectant.  
 
 
 
Base Unit, HHU, Umbilical Cable 
and Dock 
 
1. 
Determine that the HHU and Umbilical Cable are dry.  If they have not dried completely, use disposable task 
wipes to wipe down the areas until dry. 
2. 
Inspect surface for cleanliness. If not visibly clean, use one pre-saturated disinfectant wipe to completely 
preclean the surface of all gross debris. Discard wipe. 
3. 
Spray a new pre-saturated disinfectant wipe with additional disinfectant solution so that it is visibly 
saturated. Wipe the surface of the HHU with disinfectant wipe so that the entire HHU surface remains 
visibly wet for at least 1 minute. 
4. 
Saturate a large disposable swab, such as a cervical swab,  and a soft brush, with disinfectant solution. Use a 
cervical swab and brush to lightly scrub the HHU seams, button perimeters, HHU nut-housing interface, and 
cable-housing mating surface to ensure disinfectant contacts the crevices. Scrub for at least one minute with 
the swab and brush. 
5. 
After scrubbing seams, spray a new pre-saturated disinfectant wipe with additional disinfectant solution so 
that it is visibly saturated and use this wipe to vigorously rub the surface to ensure disinfectant contact for 
one minute. Discard wipe. 
6. 
Using a new pre-saturated disinfectant wipe, rub the surface of the device so that it remains visibly wet for 3 
minutes.  Repeated use of pre-saturated disinfectant wipes and disinfectant spray may be required to ensure 
the surface remains visibly wet for 3 minutes at room temperature (65F/18C).  
7. 
Discard used wipes. 
 
HHU Window 
 
 
 
 
 
1. 
Visually inspect the HHU window for visible soil or streaks. 
2. 
If the window is streaked, wipe the HHU window with pre-saturated disinfectant wipe. 
3. 
Discard wipes as they become contaminated.  
4. 
Visually inspect the HHU window for soil.  
5. 
Repeat wiping until window is not soiled.  
6. 
Disinfect window with pre-saturated disinfectant wipe. Allow the window to remain wet for 3 minutes, and 
then air dry at ambient temperature (60-80F).   
7. 
Wipe window with pre-moistened 70% IPA wipe or pre-moistened 70% IPA cotton swab applicator. Allow to 
air dry at ambient temperature  (68F/20C) 
8. 
Visually inspect the HHU window for streaks. Repeat IPA wiping until the window is dry without streaks. 
Allow to air dry at ambient temperature (68F/20C).  
Touchscreen and Bezel 
 
1. 
Caution: DO NOT spray detergent solution directly onto monitor. Rear of monitor is not protected against 
fluid intrusion. Monitor should be disinfected with pre-saturated disinfectant wipes.