VC150 Vital Signs Monitor
6-5
NIBP: NIBP on the monitor screen
Oscillometric method 
The oscillometric method of determining NIBP is accomplished by a transducer 
that measures small variations in cuff pressure resulting from the pulsatile flow 
of blood in the patient's artery while the air pressure in the cuff is varied. These 
signals are analyzed by the algorithm resulting in systolic, MAP, and diastolic 
pressure estimates calculated to correspond to either an intra-arterial or 
auscultatory reference.
NIBP on the monitor screen
The NIBP parameter field is located on the left side of the monitor screen. Blood 
pressure-related titles and pulse rate are displayed in red by default. The color of 
the unit of measurement and measurement data can be changed in 
configuration mode. The NIBP parameter field view varies depending on which 
NIBP mode is used and which part of the body is selected for the measurement.
Screen item
Description
The start icon starts the NIBP determination process 
by first inflating the cuff to the target pressure. 
Selecting the terminating icon/red inflate icon 
terminates the inflation and the NIBP process 
altogether. The NIBP mode icon above the start icon 
indicates the chosen NIBP measurement mode.
If a measurement is in progress and the cuff fills with 
air, the current pressure is displayed in digits above 
the start icon.