VC150 Vital Signs Monitor
6-7
NIBP: NIBP on the monitor screen
Alarms associated with NIBP
Physiological and technical alarms are categorized by priority level:
When an NIBP measurement invokes an alarm:
The monitor will sound an audible alarm signal.
An LED light will illuminate at the top of the monitor.
A message will be displayed in the alarm area of the monitor screen.
The NIBP value exceeding its limit will be highlighted on screen.
NOTE
Technical error conditions related to NIBP are first displayed under the DIA 
value and after 10 seconds as a low priority alarm. 
To respond quickly to an alarm, select the silence icon in upper left corner of the 
screen. Once this icon is selected, the audible alarm will remain silent for two 
minutes. The alarm will reactivate if the monitor continues to receive NIBP 
measurement information. To acknowledge an alarm, select the alarm message 
on the notification area.
PR
PR is the SpO
2
 derived pulse rate. If SpO
2
 is not being 
measured or available, PR is the NIBP derived pulse 
rate. The pulse rate label consists of the name and 
the unit of measure used. A heart symbol is 
displayed next to PR. This symbol blinks and an 
audio sound is signaled if SpO
2
 is active and a pulse 
is detected.
Limits for PR can be configured in Alarm Setup to 
trigger an alarm if the PR value does not remain 
within limits.
Screen item
Description