VC150 Vital Signs Monitor
KO00065K
Masimo SpO
2
SpO
2
 sounds
NOTE
For pediatric subjects that have limited neck space, the sensor may be 
placed on the right side of chest (dotted oval in the image above), 
underneath clavicle. The sensor should not be touching the clavicle.
3. Place sensor tape on skin.
4. Gently press on sensor tape from center outward so adhesive forms a good 
contact with patient's skin.
5. Ensure there are no skin folds or air gaps under sensor pad.
6. Remove the release liner from the anchor pad and place the anchor pad on 
patient’s side of the neck; route the sensor cable in front of patient. Do not 
place anchor pad on clothing.