Справочник Пользователя для Innokas Yhtyma Oy VC150

Скачать
Страница из 332
KO00065K
VC150 Vital Signs Monitor
8-17
Masimo SpO
2
Masimo rainbow
®
 SET
®
 SpO
2
 and special features
General Description for SpOC 
The above approximations result in the following reduced equation for oxygen 
content via the Pulse CO-Oximeter: 
SpOC (ml/dL*) = 1.31 (ml O
2
/g Hb) x SpHb (g/dL) x SpO
2
 + 0.3 ml/dL 
*When ml O
2
/g Hb is multiplied by g/dL of SpHb, the gram unit in the 
denominator of ml/g cancels the gram unit in the numerator of g/dL resulting in 
ml/dL (ml of oxygen in one dL of blood) as the unit of measure for SpOC.
Description for Carboxyhemoglobin (SpCO) 
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the 
levels of carboxyhemoglobin concentration (SpCO) in arterial blood. It relies on 
the same basic principles of pulse oximetry (spectrophotometry) to make its 
SpCO measurement. 
The measurement is obtained by placing a sensor on a patient, usually on the 
fingertip for adults and the hand or foot for infants. The sensor connects either 
directly to the Pulse CO-Oximetry instrument or through an instrument patient 
cable. 
The sensor collects signal data from the patient and sends it to the instrument. 
The instrument displays the calculated data as percentage value for the SpCO, 
which reflect blood levels of carbon monoxide bound to hemoglobin.
Successful Monitoring for SpCO
A stable SpCO reading is associated with correct sensor placement, small 
physiological changes during the measurement and acceptable levels of arterial 
perfusion in the patient’s fingertip (measurement site). Physiological changes at 
the measurement site are mainly caused by fluctuations in the oxygen 
saturation, blood concentration and perfusion.
General Description for Total Hemoglobin (SpHb)
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the 
levels of total hemoglobin (SpHb) in arterial blood. It relies on the same principles 
of pulse oximetry to make its SpHb measurement. The measurement is taken by 
a sensor capable of measuring SpHb, usually on the fingertip for adult and 
pediatric patients.
The sensor connects directly to the Pulse CO-Oximeter or with a patient cable. 
The sensor collects signal data from the patient and sends it to the instrument. 
The instrument displays the calculated data as measurement of total 