VC150 Vital Signs Monitor
11-5
Welch Allyn temperature: Welch Allyn temperature measurement
Temperature probes
The temperature probes are color-coded to indicate which probes are used for 
oral/axillary or rectal measurement sites.
Refer to the VC150 supplies and accessories document for temperature probe 
and probe cover reorder part numbers.
Proper storage of thermometer probe covers
To reduce the risk of contamination, keep the thermometer 
probe covers in their original 25-count box and store the 
box in the storage well provided on the monitor.
Welch Allyn temperature measurement
NOTES
Do not allow the tip of the temperature probe to come into contact with a 
heat source (e.g., hands or fingers) prior to taking a temperature 
measurement. If this occurs, allow 5 seconds for the probe tip to cool before 
proceeding.
Never insert a probe cover into the probe well.
Oral temperature measurement
1. Ensure that the patient has not done any of the following within 20 minutes 
prior to taking a temperature, as these may result in inability to take an 
accurate temperature:
Ingesting hot or cold liquids
Brushing  teeth
Eating
Chewing gum or mints
Smoking
Activity such as climbing stairs or running
Color-coded temperature probes
Color of the probe ejection button
Measurement site
Blue
Oral or axillary
Red