Welch Allyn temperature: Welch Allyn temperature measurement
NOTES
Do not insert the probe cover into the probe well.
Do not allow the probe tip to come into contact with the patient until the
probe is placed in the measurement site. Before this, any contact
between the probe tip and the tissue or other material may cause
inaccurate readings.
probe is placed in the measurement site. Before this, any contact
between the probe tip and the tissue or other material may cause
inaccurate readings.
8. Always hold the probe in place, maintaining tissue contact until temperature
is complete. Do not allow the patient to reposition the probe.
9. When the predictive measurement is complete, the result is captured to
patient data automatically. If you question a reading, select the snail icon to
switch to monitor mode.
switch to monitor mode.
10. When the monitor mode temperature measurement has stabilized, select
Snapshot to create a snapshot and remove the probe from the patient.
11. Eject the disposable probe cover by firmly pressing the ejection button on
the top of the probe.
CAUTION
To prevent cross-contamination, properly dispose of the probe cover
when the measurement is complete.
when the measurement is complete.
12. Place the probe in the probe well. Once you place the probe in the probe
well, the temperature values will be cleared in 1 to 30 minutes, depending
on the measurement expiration time setting.
on the measurement expiration time setting.
Axillary temperature measurement
1. Adjust clothing to access the entire axilla. Do not take an axillary
temperature through patient’s clothing. Direct probe cover to skin contact is
required.
required.
2. Check that a blue probe well and a probe with a blue ejection button are
connected to the device. If not, obtain a blue probe well and a probe with
the blue ejection button and connect these to the device as instructed in
the blue ejection button and connect these to the device as instructed in
on page 3-4. If these are already
connected, ensure that the probe is in the probe well.
3. Always place a new transparent protective temperature probe cover on the
probe before every axillary measurement:
•
Remove the probe from the probe well.
•
Insert the probe shaft into a probe cover and firmly press down until
retaining rings of the probe cover seat securely over the retaining ring
barb of the probe shaft.
retaining rings of the probe cover seat securely over the retaining ring
barb of the probe shaft.