VC150 Vital Signs Monitor
KO00065K
Maintenance: Calibration
WARNING
Do not use damaged sensors, cables, or connectors.
Calibration
User can only recalibrate the touch screen. Refer to 
on page 3-9, if necessary. No other user-lever calibrations are required. 
Cleaning
List of approved cleaning agents
Use only these cleaning agents to clean the monitor unit or the Welch Allyn 
probe well. Periodically, Innokas Medical evaluates additional cleaning agents 
for compatibility with the monitor. If your cleaning agent is not listed in this 
section, contact your Innokas Medical representative to determine if additional 
information is available.
NOTES
The touch screen surface is made of PET (polyethylene terephthalate) and 
must not be cleaned with any solvent or any alkaline cleaning agent.
If you want to use other cleaning agents than those specifically approved 
and listed in this section, you can do it at your own risk. However, bear in 
mind that using non-approved cleaning agents voids the product warranty 
on all monitor parts that may come in touch with non-approved cleaning 
agents. If you still want to use a non-approved cleaning agent, first test it on 
the corner of the touch screen, not on the whole screen. Remember to wipe 
off possible residues or stains with a cloth dampened with water.
List of approved cleaning agents for VC150 monitor and accessories 
(except Welch Allyn SureTemp
®
 Plus)
Cavicide
®
 Surface 
Cleaner/Disinfectant
Sagrotan
®
 
Alternative
Sani-Cloth
®
 HB
Sporicidin
®
Virex 256
Water (distilled)
Windex Blue
List of approved cleaning agents for Welch Allyn SureTemp
®
 Plus
Probe well
Warm water and a mild detergent solution.
Thermometry probe
70% Isopropyl alcohol 10% bleach solution and 
Cavicide
® 
or equivalent.