Справочник Пользователя для Innokas Yhtyma Oy VC150
KO00065K
VC150 Vital Signs Monitor
3-9
Getting started: Monitor on/off
Touch screen recalibration
1. To recalibrate the screen, turn off the monitor and turn it on again.
2. Wait until a blue progress bar has extended over a half of its total length.
The monitor will display the current software version and the text Touch to
recalibrate. Now press anywhere on the white screen area for three
seconds to begin the recalibration process. You can also press power button
shortly to begin recalibration.
recalibrate. Now press anywhere on the white screen area for three
seconds to begin the recalibration process. You can also press power button
shortly to begin recalibration.
NOTE
If the blue progress bar reaches the end and the black home screen
appears, you have to start the monitor again to recalibrate it.
appears, you have to start the monitor again to recalibrate it.
3. You can start the recalibration when a blinking rectangle appears in the top
left corner (A).
4. Press the rectangle to select it. Hold your finger down for a moment, if
necessary. If the software registered the selection, the rectangle moves to
the next location.
the next location.
5. Select the rectangle at each location (B, C, D and E).
NOTE
Do not slide your finger on the screen. Selections are made by pressing the
screen with your fingertip and then lifting the finger away from the screen.
screen with your fingertip and then lifting the finger away from the screen.
6. When the recalibration is complete, the home screen will appear and you
can continue using the monitor.
Verification of the touch screen recalibration
1. Select Help.
2. Close the help screen with Close.
3. Select the systolic NIBP area to jump to the NIBP screen.
4. Select the home icon to exit the NIBP screen.
If you are not able to perform the commands because the calibration is still off,
try again to recalibrate. If even the new calibration does not solve the issue,
contact service.
try again to recalibrate. If even the new calibration does not solve the issue,
contact service.