VC150 Vital Signs Monitor
KO00065K
Supplemental analysis of clinical accuracy test data: Nellcor accuracy study results
Nellcor accuracy study results
Accuracy was calculated using the root mean square difference (RMSD).
SpO
2
 Accuracy for Nellcor™ Sensors vs. CO-oximeters
SpO
2
 decade
MAX-A
MAX-N
MAX-FAST
Data Points
Arms
Data Points
Arms
Data Points
Arms
60-70
71
3.05
71
2.89
71
2.22
70-80
55
2.35
55
2.32
55
1.28
80-90
48
1.84
48
1.73
48
1.48
90-100
117
1.23
117
1.68
117
0.98