Справочник Пользователя для Innokas Yhtyma Oy VC150

Скачать
Страница из 332
D-10
VC150 Vital Signs Monitor
KO00065K
Supplemental analysis of clinical accuracy test data: Masimo sensor accuracy
M-LNCS/LNCS/LNOP - Adtx/Pdtx
M-LNCS/LNCS/LNOP - Adtx/Pdtx, measured values
Range
Arms
90-100%
1.64%
80-90%
1.07%
70-80%
1.55%
Overall Claimed Accuracy Value
 70-100%
± 2%
SaO
2
Error (SpO
- S
aO
2
 )