Справочник Пользователя для Innokas Yhtyma Oy VC150

Скачать
Страница из 332
D-14
VC150 Vital Signs Monitor
KO00065K
Supplemental analysis of clinical accuracy test data: Masimo sensor accuracy
Rainbow DCI and DCIP Sensors
The table below shows Arms (Accuracy Root Mean Square) values measured 
using the rainbow DCI and DCIP sensors with Masimo SET Oximetry Technology 
in a clinical study. For accuracy in the range of 70% - 100% please se
Rainbow DCI and DCIP, measured values
Range
Arms
75-80%
1.56%
70-75%
1.67%
65-70%
1.86%
60-65%
2.13%
Error (SpO
- SaO
2
 )
SaO
2