Справочник Пользователя для Innokas Yhtyma Oy VC150

Скачать
Страница из 332
D-12
VC150 Vital Signs Monitor
KO00065K
Supplemental analysis of clinical accuracy test data: Masimo sensor accuracy
M-LNCS/LNCS/LNOP DCI and DCIP
Table information provides Arms values measured using the M-LNCS/LNCS/
LNOP DCI and DCIP sensors with Masimo SET Oximetry Technology in a clinical 
study.
SaO
2
 versus error (SpO
2
 - SaO
2
) with linear regression fit and upper 95% and 
lower 95% limits of agreement.
M-LNCS/LNCS/LNOP - DCI and DCIP, measured values
Range
Arms
90-100%
0.6%
80-90%
0.5%
70-80%
0.7%
Overall Claimed Accuracy Value
 70-100%
± 2%
SaO
2
Error (SpO
- SaO
 )