Troubleshooting
Error code Message
Cause
Remedial action
E0179
Temperature below limit
1. Switch off the
electrosurgical generator
and wait until operating
temperature is reached.
2. If the problem persists,
contact Olympus.
The minimum operating
temperature of the
electrosurgical generator
has fallen below the limit.
Switch off the electrosurgical
generator and wait until
the specified operating
temperature is reached.
Refer to the section
“Specifications for the
CELON ELITE ESG-200” on
page 90.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0180
Temperature above limit
1. Switch off the
electrosurgical generator
and wait until it has
cooled down.
2. If the problem persists,
contact Olympus.
The maximum operating
temperature of the
electrosurgical generator
has been exceeded.
Switch off the electrosurgical
generator and let it cool
down until the specified
operating temperature is
reached. Refer to the section
“Specifications for the
CELON ELITE ESG-200” on
page 90.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0187
Increased HF leakage
current
1. Check if an instrument,
the neutral electrode or
patient is unintentionally
grounded.
2. If the problem persists,
contact Olympus.
The high frequency leakage
current has exceeded
the limit of 150 mA for
monopolar applications
or 100 mA for bipolar
applications.
Check if the electrosurgical
generator or the patient is
grounded unintentionally.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0188
Excessive HF leakage
current
1. Check if an instrument,
the neutral electrode or
patient is unintentionally
grounded.
2. If the problem persists,
contact Olympus.
The high frequency leakage
current has exceeded
the limit of 300 mA for
monopolar applications
or 200 mA for bipolar
applications.
Check if the electrosurgical
generator or the patient is
grounded unintentionally.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0202
Insufficient neutral
electrode contact
1. Check the connection and
attachment of the neutral
electrode.
2. If the problem persists,
attach a new neutral
electrode.
3. If the problem persists,
contact Olympus.
The contact resistance of
the neutral electrode is too
high or the neutral electrode
is not connected.
Check the connection and
attachment of the neutral
electrode.
Malfunction of the neutral
electrode or its connection
cable.
Replace the neutral
electrode or its connection
cable.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus