Troubleshooting
Error code Message
Cause
Remedial action
E0179
Temperature below limit
1. Switch off the 
electrosurgical generator 
and wait until operating 
temperature is reached.
2. If the problem persists, 
contact Olympus.
The minimum operating 
temperature of the 
electrosurgical generator 
has fallen below the limit.
Switch off the electrosurgical 
generator and wait until 
the specified operating 
temperature is reached. 
Refer to the section 
“Specifications for the 
CELON ELITE ESG-200” on 
page 90.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0180
Temperature above limit
1. Switch off the 
electrosurgical generator 
and wait until it has 
cooled down.
2. If the problem persists, 
contact Olympus.
The maximum operating 
temperature of the 
electrosurgical generator 
has been exceeded.
Switch off the electrosurgical 
generator and let it cool 
down until the specified 
operating temperature is 
reached. Refer to the section 
“Specifications for the 
CELON ELITE ESG-200” on 
page 90.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0187
Increased HF leakage 
current
1. Check if an instrument, 
the neutral electrode or 
patient is unintentionally 
grounded.
2. If the problem persists, 
contact Olympus.
The high frequency leakage 
current has exceeded 
the limit of 150 mA for 
monopolar applications 
or 100 mA for bipolar 
applications.
Check if the electrosurgical 
generator or the patient is 
grounded unintentionally.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0188
Excessive HF leakage 
current
1. Check if an instrument, 
the neutral electrode or 
patient is unintentionally 
grounded.
2. If the problem persists, 
contact Olympus.
The high frequency leakage 
current has exceeded 
the limit of 300 mA for 
monopolar applications 
or 200 mA for bipolar 
applications.
Check if the electrosurgical 
generator or the patient is 
grounded unintentionally.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus.
E0202
Insufficient neutral 
electrode contact
1. Check the connection and 
attachment of the neutral 
electrode.
2. If the problem persists, 
attach a new neutral 
electrode.
3. If the problem persists, 
contact Olympus.
The contact resistance of 
the neutral electrode is too 
high or the neutral electrode 
is not connected.
Check the connection and 
attachment of the neutral 
electrode.
Malfunction of the neutral 
electrode or its connection 
cable.
Replace the neutral 
electrode or its connection 
cable.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator.
Contact Olympus