Справочник Пользователя для Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S

Модели
MSO1074Z-S
Скачать
Страница из 172
RIGOL 
2-86 
  DS1000Z Programming Guide 
:TRIGger:EDGe:LEVel 
Syntax 
:TRIGger:EDGe:LEVel <level> 
:TRIGger:EDGe:LEVel?
 
Description 
Set or query the trigger level in edge trigger. The unit is the same as the current 
amplitude unit. 
 
Parameter 
Name 
Type  Range 
Default 
<level> 
Real 
(± 5 × VerticalScale from the screen center)  -OFFSet  0 
 
Explanation 
For VerticalScale, refer to the 
 command. For OFFSet, refer to 
the 
 command. 
 
Return 
Format 
The query returns the trigger level in scientific notation.   
Example 
:TRIGger:EDGe:LEVel 0.16 
Related 
Command