Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S 用户手册
产品代码
MSO1074Z-S
RIGOL
2-86
DS1000Z Programming Guide
:TRIGger:EDGe:LEVel
Syntax
:TRIGger:EDGe:LEVel <level>
:TRIGger:EDGe:LEVel?
:TRIGger:EDGe:LEVel?
Description
Set or query the trigger level in edge trigger. The unit is the same as the current
amplitude unit.
Parameter
Name
Type Range
Default
<level>
Real
(± 5 × VerticalScale from the screen center) -OFFSet 0
Explanation
Return
Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:EDGe:LEVel 0.16
Related
Command