RIGOL 
DS1000Z Programming Guide 
2-97
 
:TRIGger:VIDeo:STANdard 
Syntax 
:TRIGger:VIDeo:STANdard <standard> 
:TRIGger:VIDeo:STANdard? 
Description 
Set or query the video standard in video trigger.   
Parameter 
Name 
Type 
Range 
Default 
<standard> 
Discrete  {PALSecam|NTSC|480P|576P} 
NTSC 
 
Explanation 
  PALSecam:   
PAL: the frame frequency is 25 frames per second. The TV sweep line is 625 with the 
odd field goes first and the even field follows behind. 
Secam: the frame frequency is 25 frames per second. The sweep line is 625 with 
interlacing sweep.
 
  NTSC: the field frequency is 60 fields per second and the frame frequency is 30 
frames per second. The TV sweep line is 525 with the even field goes first and the 
odd field follows behind.
 
  480P: the frame frequency is 60 frames per second; the TV sweep line is 525; 
line-by-line sweep; the line frequency is 31.5 kHz.
 
  576P: the frame frequency is 60 frames per second; the TV sweep line is 625; 
line-by-line sweep.
 
Return 
Format 
The query returns PALS, NTSC, 480P or 576P.   
Example 
:TRIGger:VIDeo:STANdard NTSC 
Related 
Command
 
 
 
:TRIGger:VIDeo:LEVel 
Syntax 
:TRIGger:VIDeo:LEVel <level> 
:TRIGger:VIDeo:LEVel? 
Description 
Set or query the trigger level in video trigger. The unit is the same as the current 
amplitude unit. 
 
Parameter 
Name 
Type  Range 
Default 
<level> 
Real 
(± 5 × VerticalScale from the screen center)  -OFFSet  0 
 
Explanation 
  For VerticalScale, refer to 
. 
 
  For OFFSet, refer to 
 
Return 
Format 
The query returns the trigger level in scientific notation.   
Example 
:TRIGger:VIDeo:LEVel 0.16   /*Set the trigger level to 160mV*/ 
:TRIGger:VIDeo:LEVel   /*The query returns 1.600000e-01*/ 
Related 
Command